IBM M/02-212 Manual do Utilizador Página 59

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 65
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 58
Brookfield Engineering Labs., Inc. Page 59 M/02-212
Zero Calibration Error
Cause: An impermissible high value was measured during carrying out of zero point calibration of
the Rheometer
Message:
What to do:
Push key.
Check if the measuring system has been really removed from the Rheometer.
Retry zero point calibration.
If the error is presented again, that means that there is a fault in the measuring instrument
BREAK: MOMMAX
Cause: Maximum torque has been exceeded during carrying out of a measurement.
a) You have selected inappropriate measuring range load for the measuring medium.
b) Rotation bob distorted in the measuring cup.
Message:
What to do:
Push key.
Check if the measuring system is fixed at the Rheometer properly. If so, select for this
measurement smaller shear rates or speed or use a measuring system with higher shear
stress factor.
Error #3
Zero cal. error!
Please retry cal.
<OK> cont. <ST> stop
BREAK: torque max!
Single mode
total MP: 10
total time: 60
ST
OK
Vista de página 58
1 2 ... 54 55 56 57 58 59 60 61 62 63 64 65

Comentários a estes Manuais

Sem comentários